Key Features
Easily rotatable display, 90° increments
Easy local adjustments using 3 pushbuttons
Flow measurement with SQRT function
Min. / Max. function records low and high pressure events
Compact, rugged construction
“Loop check” allows easy display of user-defined units and verification without applying pressure
Bright backlit 4-digit LCD display
Markets & Applications
Food and Beverage
Filtration Monitoring
Pressure and Non-Pressurized Tank Levels
Dampening
Digital by firmware setting
Process Connection Style
Threaded
Output Signal
4-20 mA (2-wire)
Accuracy
0.5% of span (URL)
Stability
≤±0.25% span/year at reference conditions
Ranges
Unidirectional: 10 ... 1000 mbar / 4 ... 400 in. H2O
Bidirectional: ±10 ... ±500 mbar / ±4 ... ±200 in. H2O
Pressure Type
Differential pressure
Wetted Parts Material
Ceramics
Stainless steel 316L (1.4404)
Viton®
Static Pressure
20 bar / 300 psi
Ingress Protection
IP66 / NEMA 4
Display
4 digit LCD with LED backlight
Case or Body Material
Epoxy coated aluminium
Process Connection Location
Lower
Mounting
Surface
Stem
Electrical Termination
Stainless steel cable gland
1/2 NPT female
Equipped with a highly reliable and accurate sensing element, featuring the patented Ashcroft® Si-Glass™ sensor, Ashcroft offers a range of differential pressure transmitters specifically designed and developed for use in the cleanroom industry. Using an ultra-thin single crystal diaphragm, Ashcroft differential pressure transmitters offer inherent sensor repeatability and stability, making them stand out as highly accurate, long-term stable and reliable measurement instruments for cleanroom pressure monitoring.
The silicon diaphragm sensor is free of adhesives or other organic materials that are commonly used in differential pressure measurement technology and contribute to a drift in the measured value acquisition or a mechanical reduction in the measurement accuracy.
Ashcroft differential pressure transmitters incorporate the TruAccuracy™ specification. Ashcroft’s accuracy claim is based solely on terminal endpoint methodology and not on statistically derived accuracy claims’.
TruAccuracy™ means that Ashcroft differential pressure transmitters include zero offset, non-linearity, hysteresis, non-repeatability, zero offset and span adjustment error in their accuracy statement.
The user thus has a ready-to-install device at his disposal that does not require any initial calibration after installation. This considerably reduces the time and costs involved.
Other commercially available measuring devices usually do not include information on zero point deviation and span error in their accuracy specifications, so that these measurement errors (up to ±1.00%) must be added to the accuracy in order to achieve reliable performance characteristics.