Key Features
9 engineering units: 1 field-programmable
Large 5 digit LCD display
Min. / Max. function records low and high pressure events
Simple Push-button menu navigation
20 segment Bar graph display
Wide variety of pressure connection thread options
Protective boot (optional)
Customisable keypad (optional)
Markets & Applications
Pharmaceuticals
Pumps and Compressors
Food and Beverage
Machine Automation
Torque Measurement
Hydraulics and Pneumatics
Material Strength Testing
Weighing Systems
Ranges
Compound
1 ... 1600 bar / 15 ... 25000 psi
Vacuum
Pressure Type
Gauge pressure
Wetted Parts Material
Stainless steel 316L (1.4404)
Sensor stainless steel 17-4PH
Ingress Protection
IP67
Case or Body Material
Polycarbonate/ABS
Keypad Polycarbonate
Process Connection Location
Lower
Mounting
Stem
Dampening
Update rate by firmware
Throttle screw
Case Style
Open front case
Process Connection Style
Threaded
Dial Size
2 1/2"
Display
Full 5 digital LCD
Accuracy
0.25% of span (ASME B40.7 Grade 3A)
0.5% of span (ASME B40.7 Grade 2A)
Equipped with a highly reliable and accurate sensing element, featuring the patented Ashcroft® Si-Glass™ sensor, Ashcroft offers a range of differential pressure transmitters specifically designed and developed for use in the cleanroom industry. Using an ultra-thin single crystal diaphragm, Ashcroft differential pressure transmitters offer inherent sensor repeatability and stability, making them stand out as highly accurate, long-term stable and reliable measurement instruments for cleanroom pressure monitoring.
The silicon diaphragm sensor is free of adhesives or other organic materials that are commonly used in differential pressure measurement technology and contribute to a drift in the measured value acquisition or a mechanical reduction in the measurement accuracy.
Ashcroft differential pressure transmitters incorporate the TruAccuracy™ specification. Ashcroft’s accuracy claim is based solely on terminal endpoint methodology and not on statistically derived accuracy claims’.
TruAccuracy™ means that Ashcroft differential pressure transmitters include zero offset, non-linearity, hysteresis, non-repeatability, zero offset and span adjustment error in their accuracy statement.
The user thus has a ready-to-install device at his disposal that does not require any initial calibration after installation. This considerably reduces the time and costs involved.
Other commercially available measuring devices usually do not include information on zero point deviation and span error in their accuracy specifications, so that these measurement errors (up to ±1.00%) must be added to the accuracy in order to achieve reliable performance characteristics.