Key Features
FM, CSA and ATEX Intrinsically safe versions for Hazardous Locations (optional)
Dual sensor module bay for interchangeable modules
Data logging capability up to 16,000 records
USB communications port
SD memory card slot
Standard firmware features – min./max., tare, programmable damping, percent function, and switch test
Built in Power supply to power device under test
Large LCD display with backlight
Interchangeable pressure modules with ranges from 60 Pa up to 700 bar
Markets & Applications
HVAC-R
Chemical and Petrochemical
Energy
Food and Beverage
Biotech
Oil and Gas
Machine Automation
Test and Measurement
Calibration Laboratories
Hydrogen and Technical Gases
Pharmaceuticals
Hospitals and Medical Research
Pumps and Compressors
Temperature Rating
-20 ... 49°C
Ranges
Vacuum
Compound
Pt100: -200 ... 550 °C
0 ... 30 VDC
Type K: –200 ... 1000°C
0 ... 20 mA
60 Pa ... 700 bar
Cable Material
High Impact PC-ABS
Wetted Parts Material
Aluminum, Titanium, PBT, Buna, Glass, Gold, Silicone Rubber, Silicon, Silicone RTV & Brass. NOT FOR USE WITH LIQUIDS
Stainless steel 316L (1.4404)
Ingress Protection
IP65
Accuracy
Low pressure modules: ±0.1%, ±0.06, ±0.07% of span
Medium/high pressure modules ±0.025%, ±0.05%, ±0.1% of span
Pressure Type
Gauge pressure
Absolute pressure
Differential pressure
Display
2.5 ̋ (W) x 1.5 ̋ (H) displays 3 simultaneous measurements, Back light, Bar Graph, Flip screen capability
Electrical Termination
4 mm banana jacks
Equipped with a highly reliable and accurate sensing element, featuring the patented Ashcroft® Si-Glass™ sensor, Ashcroft offers a range of differential pressure transmitters specifically designed and developed for use in the cleanroom industry. Using an ultra-thin single crystal diaphragm, Ashcroft differential pressure transmitters offer inherent sensor repeatability and stability, making them stand out as highly accurate, long-term stable and reliable measurement instruments for cleanroom pressure monitoring.
The silicon diaphragm sensor is free of adhesives or other organic materials that are commonly used in differential pressure measurement technology and contribute to a drift in the measured value acquisition or a mechanical reduction in the measurement accuracy.
Ashcroft differential pressure transmitters incorporate the TruAccuracy™ specification. Ashcroft’s accuracy claim is based solely on terminal endpoint methodology and not on statistically derived accuracy claims’.
TruAccuracy™ means that Ashcroft differential pressure transmitters include zero offset, non-linearity, hysteresis, non-repeatability, zero offset and span adjustment error in their accuracy statement.
The user thus has a ready-to-install device at his disposal that does not require any initial calibration after installation. This considerably reduces the time and costs involved.
Other commercially available measuring devices usually do not include information on zero point deviation and span error in their accuracy specifications, so that these measurement errors (up to ±1.00%) must be added to the accuracy in order to achieve reliable performance characteristics.